10. Appendices > 10.9. Advanced maintenance procedures
10.9.1. Adding water to DISCfill
1.
Unscrew the white cap from DISCfill.
2.
Fill DISCfill with distilled water. DISCfill holds approximately 1.2 liters (0.3 gallons) of water.
3.
Replace the cap.
Typically, DISCfill does not need filling between planned maintenance. If DISCfill needs to be filled
more frequently, contact Service.
10.9.2. Removing air from the filling hose
If you suspect that there is air in the filling hose that prevents the water flow, remove the air by the
following method. You need the draining hose and a container for water (a washbowl, for example).
1.
Leave DISCfill on top of the ultrasound window.
2.
Disconnect the filling hose from the Sonalleve Table Top. Leave the filling hose connected to
DISCfill.
3.
Position the container at the side of the Sonalleve Table Top, below the level of DISC.
4.
Place the open end of the draining hose into the container.
5.
Connect the filling hose into the female hose connector of the draining hose.
6.
Position the hoses so that at no point between DISCfill and the container the water flow needs to
turn upwards.
7.
Let some water flow through the hoses.
8.
Disconnect the filling hose from the draining hose.