2. Safety > 2.5. Emergency Stop and Safety Devices

2.5. Emergency Stop and Safety Devices

There are three abort mechanisms in the Sonalleve MR-HIFU system:
Patient Emergency Stop Button (PESB) for the patient.
A Stop button in the Safety Device for the user.
An abort button in the user interface.
WARNING
Use the Safety Device Stop button or PESB immediately in a hazardous situation. Use the
software abort button to stop the sonication in situations that do not present any hazard (if,
for example, the treatment is deemed sufficient before the planned treatment time is
reached).

2.5.1. Patient Emergency Stop Button (PESB)

The patient can stop the treatment at any point by pressing the Patient Emergency Stop Button
(PESB). Pressing the PESB terminates the sonication immediately, although scanning of the
temperature maps continues. As the patient presses the button, an indicator light is switched on in
the device, there is a beep sound in the MR room and in the operator's Safety Device, and the
Sonalleve Console displays an error. Always instruct the patient about the correct usage of the
PESB prior to the start of the treatment. If the patient has stopped the treatment because of pain,
analyze the cause of pain before continuing the treatment.
When the patient presses the PESB, the red indicator light in the Safety Device is switched on. No
further sonications are possible, before the user presses the green Enable button in the Safety
Device.
During pre-treatment scanning, and when sonication is not on, the PESB functions like the MR
nurse call button: it gives an audio signal but does not stop scanning.
WARNING
Do not use the standard MR Nurse Call button with the Sonalleve MR-HIFU system. Use the
Patient Emergency Stop Button (PESB) instead.
Instructions for Use
109745C2 / 02-2022
18 (192)