6. Workflow > 6.7. Preparing the system for sonication
WARNING
QuickCover only provides protection for the superficial tissue layer immediately behind
the cover strip.
Using protective cover strips side-by-side may lead to distorted heating due to the large
area of blocked ultrasound beam.
Using protective cover strips side-by-side may lead to transducer damage due to the
large intensity of reflected ultrasound.
Air bubbles trapped on the skin at an edge of a protective cover strip may lead to a local
skin burn. This risk is especially pronounced if protective cover strips are used so that
they overlap with each other.
Storing QuickCover strips under a weight may lead to compression and a permanent
loss of protective capabilities.

6.7. Preparing the system for sonication

6.7.1. Hardware Initialization

The treatment cannot be started until the hardware initialization is successful. Planning is possible
during hardware initialization.
NOTICE
The MR scanner cannot export images to the hospital network once Network Isolation is in
progress. Before starting the Hardware Initialization, make sure that there are no image export
tasks running or queued on the MR workstation.
Press the green Enable button on the Safety Device to enable the ultrasound generators.
1.
2.
Perform the hardware tests. If any of the tests fail, check the message log for problems in the
system. Retry until the tests succeed.
Instructions for Use
109745C2 / 02-2022
Fig. 83: The correct use of QuickCover.