The Discfill Tool; Technical Specifications - Profound Sonalleve MR-HIFU Instructions For Use Manual

Table of Contents

Advertisement

3. System overview > 3.4. Technical specifications

3.3.1. The DISCfill tool

The DISCfill tool is provided for the possible need to refill DISC between planned maintenance.
DISCfill is a container, from which DISC can be filled. There is an indicator on DISCfill for when a
refill of DISC is required.
Always use DISCfill before a HIFU treatment to check the water level in DISC and refill DISC
if the water level is too low.
For connecting DISCfill to DISC, a filling hose is supplied with DISCfill. There is a hose connector
at both ends of the filling hose. There is also a draining hose, with one end open.

3.4. Technical specifications

Tab. 5: Key treatment-related technical specifications for the Sonalleve MR-HIFU system.
Feature
Temperature measurement update
interval
Temperature change measurement
accuracy
Treatment point positioning
accuracy
Instructions for Use
109745C2 / 02-2022
Fig. 12: DISCfill (1), The filling hose (2), The draining hose (3)
Specification
3.5 seconds or less
±5 °C or better, relative accuracy (standard deviation) is
±1 °C or better
3 mm or better
41 (192)