9.2.3 Overview of adjustment parameters in MyModes

CAUTION
Improper use of the setting parameters in the MyModes
Falling due to unexpected product behaviour because of changed damping behaviour.
► Instruct the patient regarding the functionality and adjustment options for all parameters of the MyModes.
INFORMATION
When mute mode (silent mode) is activated, no beep and vibration signals are generated.
The parameters in the MyModes describe the static behaviour of the prosthesis for a specific movement pattern
such as cycling. Damping behaviour is not automatically controlled and adjusted in MyModes.
Parameter
Basic flex.
Gain
Basic ext.
Locking angle
Pitch
Volume
42
Adjustment soft­
Setting range,
ware range
app
Deactivated
0 - deactivated
Activated
1 - activated
1000 Hz
1000 Hz
4000 Hz
4000 Hz
0 – 4
0 – 4
Adjustment soft­
Setting range,
ware range
app
0–200
+/- 20
0–100
+/- 10
0–60
+/- 20
0–90
+/- 10
1000 Hz–4000 Hz 1000 Hz–4000 Hz Pitch of beep signal for confirmation tones
0–4
0–4
Meaning
Activating/deactivating the stair and obstacle
function. For switching with the Cockpit app,
this function has to be activated in the adjust­
ment software. Further information (see
page 38).
Pitch of beep signal for confirmation tones
Volume of beep signal for confirmation tones
(e.g. when checking the charge level, switch­
ing MyModes). The "0" setting deactivates
the audible feedback signals. However,
warning signals in case of errors are still gen­
erated.
Meaning
Level of flexion resistance at the start of flex­
ing the knee joint
Increase in flexion resistance (starting with
the parameter "Basic flex.") when flexing the
knee joint. The knee joint locks at a certain
flexion angle, which depends on the setting
of the parameters "Basic flex." and "Gain".
Level of extension resistance
Angle up to which the knee joint can be
extended.
Information: If this parameter is >0, the knee
joint is locked in a flexed position in the exten­
sion direction. To unlock it, unload the pros­
thesis and tilt it back for at least 1.5 seconds.
This makes extension of the joint possible
independently of the settings for the paramet­
ers "Basic ext." and "Locking angle". This
may be necessary to switch to basic mode
using a movement pattern.
Volume of beep signal for confirmation tones
(e.g. when checking the charge level, switch­
ing MyModes). The "0" setting deactivates
the audible feedback signals. However,
warning signals are still generated if errors
occur.