PC" Bluetooth adapter. The installation and use of the "60X5 BionicLink PC" adapter are described in the instruc­
tions for use supplied with the adapter.

7.1.1.3 Preparing the product to connect to the adjustment software

If the product does not emit any signals when querying the charge level (see page 27), the battery is drained or the
product is switched off.
Switching on the product
1) Connect the power supply with battery charger to the wall socket.
2) Connect the battery charger to the product.
3) Wait for feedback signals.
4) Disconnect the battery charger from the product.
→ After feedback signals are emitted (self test), the product is switched on.
Switching on Bluetooth
Upon delivery the Bluetooth function of the prosthesis is switched on.
The Bluetooth function can be switched off using the Cockpit app or the adjustment software. When the Bluetooth
function is switched off, it is only turned on temporarily for 2 minutes after connecting/disconnecting the battery
charger and is then turned off again automatically. When a connection with the PC is active (the
lit up), the Bluetooth function is not switched off automatically.

7.1.2 Removing the Protective Cover

1) Open the closures.
2) Spread open the Protective Cover and remove it from the knee joint.
3) Fasten the two closures after removing the Protective Cover.

7.1.3 Shortening the Tube Adapter

CAUTION
Incorrect processing of tube
Falling due to damage to the tube.
► Do not clamp the tube into a vice.
► For shortening the tube, use only a tube cutter.
CAUTION
Damage to the cable while shortening the tube adapter
Falling due to unexpected product behaviour as the result of switching into safety mode.
► When shortening the tube adapter, make sure the cable does not get damaged.
1) Determine the required length of the tube adapter using the configuration assistant in the adjustment software.
2) Shorten the tube adapter to the determined value with the 719R3 tube cutter.
3) Store the tube adapter cable in the tube adapter. If this is not possible, the cable must be protected against
damage.
4) Use a file (cut 2 (medium), e.g. 715H1=2 recommended) to file the cut edge smooth. Be careful of the tube
adapter cable.
NOTICE! When filing or deburring, make sure that no metal shavings can get into the plug of the tube
adapter cable.
5) Chamfer the outside with a file.
6) Smooth the inside and outside of the cut edge with sandpaper (recommended grit 120).

7.1.4 Installing the Tube Adapter

CAUTION
Improper assembly of the screw connections
Falling due to breakage or loosening of the screw connections.
► Clean the threads before every installation.
► Apply the specified tightening torque values for installation (see the section "Technical data" see page 51).
► Observe the instructions for securing the screw connections and the use of the correct length.
Genium X3 - Bionic Prosthetic System 3B5-X3/3B5-X3=ST,Ge­
nium X3 3B5-2/3B5-2=ST
Preparing the product for use
symbol is
29