Appendices
Beep signal
5 x long
10 x long
30 x long
–
Error while charging the product
LED on
Status LED
power sup
on battery
ply
charger
Status LED Charging status indicat
The
LED
ring is lit in
weak purple
56
Vibration signal
5 x long repeated every
60 seconds
10 x long
1 x long, 1 x short
repeated every 3 seconds
Continuous
Error
Country-specific plug adapter not fully
engaged on power supply
Non-functional socket
Defective power supply
No connection between battery charger
and power supply
Defective battery charger
or (5 LEDs)
No LED is lit
Result
Medium
error
(see
page 48)
e.g. a sensor is not oper
ational
Battery charge level 5%
After the beep and vibra
tion signals, the product
switches to empty battery
mode and then switches
off.
Severe error/indication
of safety mode activa
tion (see page 48)
e.g. one or more sensors
are not operational.
Total failure
Electronic
control
no
longer possible. Safety
mode active or undeter
mined
valve
state.
Undetermined
product
behaviour.
Resolution
Check whether the country-specific plug
adapter is fully engaged on the power
supply.
Check socket with another electric appli
ance.
The battery charger and power supply
must be inspected by an authorised
Ottobock Service Centre.
Check whether the charging cable plug is
fully engaged on the battery charger.
The battery charger and power supply
must be inspected by an authorised
Ottobock Service Centre.
Error
Distance between battery
charger and receiver of the
charging unit on the pros
thesis too great. If the dis
tance is more than 2 mm,
the prosthesis cannot be
charged.
Genium X3 - Bionic Prosthetic System 3B5-X3/3B5-X3=ST,Ge
Required action
Walking possible with restric
tions. Please note the change
in flexion resistance.
The product must be inspec
ted immediately by an author
ised Ottobock Service Center.
Charge the battery.
Attempt to reset this error by
connecting/disconnecting the
battery charger.
If the error persists, use of the
product
is prohibited. The
product must be inspected by
an authorised Ottobock Ser
vice Center.
Attempt to reset this error by
connecting/disconnecting the
battery charger.
If the error persists, use of the
product
is prohibited. The
product must be inspected by
an authorised Ottobock Ser
vice Center.
Resolution
Reduce distance between
battery charger and receiver
of the charging unit.
nium X3 3B5-2/3B5-2=ST