Technical data
Product
Minimum knee rotation point-ground measurement for
use of 2R19
Minimum distal system height with 2R19 tube adapter
Maximum distal system height with 2R19 tube adapter
Range of Bluetooth connection to PC
Range of Bluetooth connection to mobile device
Maximum possible flexion angle
Maximum possible flexion angle with preinstalled 4H102
flexion stop
Maximum possible flexion angle with 4H101 flexion stop 127.5°
Maximum possible flexion angle with 4H104 flexion stop 112.5°
Maximum insertion depth of the tube adapter in the knee
joint
Weight of the prosthesis without tube adapter and Pro­
tective Cover
Tube adapter
Reference number
Weight
Material
Max. body weight
Protection rating
Water resistance
Lifetime
Prosthesis battery
Battery type
Charging cycles (charging and discharging cycles)
after which at least 80% of the original battery capacity
remains available
Charge level after 1 hour charging time
Charge level after 2 hours charging time
Charge level after 4 hours charging time
Charge level after 8 hours charging time
Product behaviour during the charging process
Operating time of the prosthesis with new, fully charged
battery at room temperature
Power supply
Reference number
Storage and transport in original packaging
Storage and transport without packaging
Operation
Input voltage
Mains frequency
Output voltage
Battery charger
Reference number
Storage and transport in original packaging
52
318 mm
310 mm
514 mm
max. 10 m
max. 10 m
135°
120°
70 mm
approx. 1710 g
2R19
190 g–300 g
Aluminium
125 kg
IP66/IP68
Maximum water depth: 3 m
Maximum time: 1 hour
Waterproof, resistant to corrosion, resistant to penetra­
tion from water jets
6 years
Li-Ion
500
30 %
50 %
80 %
Fully charged
The product is non-functional
Approx. 5 days with average use
757L16*
-40 °C/-40 °F to +70 °C/+158 °F
-40 °C/-40 °F to +70 °C/+158 °F
10 % to 93 % relative humidity, non-condensing
0 °C/+32 °F to +40 °C/+104 °F
max. 90 % relative humidity, non-condensing
100 V~ to 240 V~
50 Hz to 60 Hz
12 V
4E60*
-25 °C to 70 °C/-13 °F to 158 °F
Genium X3 - Bionic Prosthetic System 3B5-X3/3B5-X3=ST,Ge­
nium X3 3B5-2/3B5-2=ST