Display Of The Current Charge Level; Display Of Battery Charge Level Without Additional Devices; Display Of The Current Charge Level Using The Cockpit App - Otto Bock Genium X3 3B5-3 Instructions For Use Manual

Hide thumbs Also See for Genium X3 3B5-3:
Table of Contents

Advertisement

Available languages

Available languages

INFORMATION
Charging the product without tube adapter
When the inductive charger is disconnected from the knee joint without the tube adapter, five beep and vibration
signals are output. In order to avoid this, perform the following steps before disconnecting the charger:
► Hold the joint upright and flex the knee head 90°.
► Wait 5 seconds. The joint switches to energy saving mode.
→ The tube adapter can be disconnected without generating a warning signal.

6.3 Display of the current charge level

6.3.1 Display of battery charge level without additional devices

INFORMATION
The charge level cannot be displayed during the charging process, e.g. by turning the prosthesis over. The
product is in charging mode.
Beep signal
5x short
4x short
3x short
2x short
1x short
INFORMATION
If the Volume parameter is set to '0' in the Cockpit app (see page 91) or if mute mode (silent mode) is activated,
there are no beep signals.

6.3.2 Display of the current charge level using the Cockpit app

Once the Cockpit app has been started, the current charge level is displayed in the bottom line of the screen:
Genium X3 3B5-3/3B5-3=ST
1) Connect the inductive charger to the receiver of the charging unit on the rear of
the product. The charger is held in place by a magnet.
→ The LED ring on the rear of the charger pulsates purple (4-second cycle).
→ If the LED ring lights up in a different colour, this indicates an error (see
page 106).
2) The charging process starts.
→ Once the product battery is fully charged, all LEDs on the side of the battery
charger light up.
3) After the charging process is complete, hold the prosthesis still and remove the
inductive charger from the receiver.
→ A self-test is performed. The joint is operational only after corresponding
feedback (see page 109).
1) Turn the prosthesis by 180° (the sole of the foot must face up).
2) Hold still for 2 seconds and wait for beep signals.
Battery charge level
more than 80%
60% to 80%
40% to 60%
20% to 40%
less than 20%
Charging the battery
79

Advertisement

Chapters

Table of Contents
loading

This manual is also suitable for:

Genium x3 3b5-3-st