Maintenance

800-cycle or biannual maintenance

SEQUENCE OF PROCEDURES TO CLEAN THE WATER TANK
To clean the water tank, proceed as follows:
 
"Prepare the sterilizer for cleaning the water tank" below "Prepare the sterilizer for cleaning the water tank" below.
1
 
"Access to the water tank" on the next page" "Access to the water tank" on the next page.
2
 
"Clean the water tank" on page  72
3
PREPARE THE STERILIZER FOR CLEANING THE WATER TANK
Proceed as follows:
 
Switch OFF the sterilizer and disconnect the mains cable.
1
 
Completely drain both water tanks (see "Draining the used and clean water tank"
2
on page  76).
 
To drain the detergent solution during the subsequent cleaning, leave the drain
3
tube attached to the drain connection of the tank you want to clean.