Maintenance
Ordinary maintenance
MAINTENANCE BY THE USER
1
1
Frequency
Cycles
Operation
Cleaning the door gasket and the chamber face
Monthly
50
side. See "Cleaning the door gasket and the
chamber face side" on page 66
Clean the chamber, trays and the rack. See
"Cleaning the chamber and the chamber
accessories" on page 67
Cleaning the chamber filters. See "Cleaning the
chamber filters" on page 68.
Cleaning the external surfaces of the sterilizer.
See "Cleaning the external surfaces of the
sterilizer" on page 68
6 month
800
Clean the clean water tank. See "800-cycle or
biannual maintenance" on page 70.
2
2
Replace the bacteriological filter. See "400-cycle
Yearly
400
maintenance" on page 69.
2
2
Replace the door gasket. See "800-cycle
Yearly
800
maintenance" on page 73.
5 years
4000
General check and service. See "4000 cycle or
five-year maintenance" on page 75.
1
: whichever occurs first.
2
: even if the maximum cycle number is not reached, it is recommended to
replace the consumable parts every year, or if they appear worn or
damaged, or if the filters are clogged or discolored.
64
EXPIRED MAINTENANCE
The sterilizer monitors the wear of consumables by counting the
number of cycles executed since the last replacement.
When the number of cycles is close to the maximum, a pre-alert
about the concerned consumable is displayed. Please check that
you have the requested spare part available, buy one if not. When
the maximum number of cycles has been met, a message to replace
the consumable will be displayed.
If you can not replace the consumable immediately, the sterilizer
will operate anyway but the message will appear again some cycles
later.
Tap
to see an animated replacement procedure.
1
When you have replaced the consumable tap
2
executed cycle counter is reset.
Lara XL | Instructions for Use | RIS-311 ENG Rev02 | 01/06/2022 | © W&H Sterilization Srl