COMMON STERILIZATION CYCLE DATA
 
 
Sterilization temperature
Sterilization pressure
Duration of the Plateau/Sterilization phase
Duration of the drying phase (ECO DRY plus mode)
Minimum duration of the drying phase (set by the user)
Load type
1
:"dental load" includes turbine and handpiece family.
Lara XL  | Instructions for Use |  RIS-311 ENG Rev02 |  01/06/2022 |  © W&H Sterilization Srl
B Universal 134
B Prion / Extended 134
134 °C (273 °F)
134 °C (273 °F)
3.03 bar
2.03 bar (g)
4'
5'-22' (ECO DRY +)
5'-22' (ECO DRY +)
22'
All unwrapped, bagged, single/double wrapped items:
 
Solid
n
 
Hollow A (narrow lumen)
n
 
Hollow B (simple hollow item)
n
 
Porous
n
Sterilization cycles
B Universal 121
121 °C (250 °F)
3.03 bar
2.04 bar
2.03 bar (g)
1.04 bar (g)
18'30''
20'30''
5'-28' (ECO DRY +)
22'
28'
Technical data
S Fast 134 (optional)
134 °C (273 °F )
3.03 bar
2.03 bar (g)
3'30''
0'-4' (ECO DRY +)
4'
Unwrapped items:
 
Solid
n
 
Hollow B (simple
n
hollow item)
1
 
Dental load
n