Sterilizer tests
INTERPRET THE TEST RESULT
Indicator
44
What happened
The entire surface of the indicator sheet has changed colour.
Certain areas of the indicator sheet have not changed colour since there was an
air pocket during the cycle due to sterilizer malfunction.
Test
passed
Yes
No
Lara XL  | Instructions for Use |  RIS-311 ENG Rev02 |  01/06/2022 |  © W&H Sterilization Srl
What to do next
-
Repeat the test. If it fails
repeatedly, call technical service.