W&H Lara RIS-311 Instructions For Use Manual page 54

Hide thumbs Also See for Lara RIS-311:
Table of Contents

Advertisement

Sterilization cycles
 
Check the cycle requirements.
2
 
Check the icon in the top left corner of the page to know the
3
ECO DRY mode assigned to the cycle.
 
Tap
:
4
 
to change the ECO DRY mode (see "Change the ECO DRY
n
mode" on page  56).
 
to disable the ECO DRY mode (see "Set the drying time" on
n
page  56).
 
If the door gasket is new, hold the door gently closed until the
5
beginning of the cycle.
 
Tap
and enter your credentials if required: the door locks. If
6
you have not set a different start time, the sterilization starts
immediately.
 
Wait the end of the sterilization. Tap
7
parameters in real time. See "View the cycle parameters" on
page  56.
 
The sterilization is completed. Tap
8
summary or tap
to view the cycle information. See "View the
cycle parameters" on page  56.
 
Tap OPEN: the door unlocks and the homepage appears.
9
54
to view the cycle
to view the cycle
 
If required, enter your credentials and confirm the load release.
10
Note : the load release is possible only if the cycle is completed
successfully.
Note: the user visually checks that pouches are intact and dry. It is even
possible to identify the user who releases the load.
Note: during the load release it is also possible to insert the chemical
indicator result (for AU-NZ version only).
   
SET THE STERILIZATION CYCLE START
You may schedule the start of the sterilization cycles at a certain
date and time (e.g., if you want to load the sterilizer in the evening
and run standard sterilization cycle early the next morning before
office hours). You can set the cycle start date and time and enable or
disable it for each cycle.
Lara XL  | Instructions for Use |  RIS-311 ENG Rev02 |  01/06/2022 |  © W&H Sterilization Srl