Technical data
WATER SUPPLY SYSTEM
Temperature
max. 35 °C (95 °F)
Pressure
min. 2 bar – max. 8.6 bar (min 29 psi - max. 124.7 psi)
Flow
min. 0.25 – max. 0.5 l/min (min. 0.066 - max. 0.132 gal/min)
POWER SUPPLY SYSTEM
Nominal voltage and Max.
200–240 V ac (±10%), 50/60 Hz, 10 A, single-phase
current
Overvoltage category
II
Suitable circuit breaker and a Ground Fault Circuit Interrupter
Protection required
(GFCI). All protection devices must be certified according to
applicable standard. A grounded connection is essential.
Communication with other
1 USB port in the front
devices
4 USB ports - 1 LAN port in the rear
Features
Fully micro-processor controlled, process evaluation system
according to EN13060.
Programmable standby mode.
Max. heat output
3000 kJ/h
INSTALLATION REQUIREMENTS
From +5 °C to +40 °C (from +41 °F to +104 °F)
Working temperature
Working relative humidity
Max. RH 80% up to 31 °C (88 °F), linearly decreasing to 50% at 40
°C (104 °F)
Lara XL | Instructions for Use | RIS-311 ENG Rev02 | 01/06/2022 | © W&H Sterilization Srl
From -20 °C to +60 °C (from -4 °F to +140 °F ) / 0–90 % (with
Storage temperature / rel.
humidity
empty tanks)
Max altitude
3000 m asl
0.6 bar (8.7 psi)
Min. atmospheric
pressure
Overall dimensions
W: 49 cm/H: 46 cm/D: 62 cm (W: 19.3''/H: 17.9''/D: 24.4'')
W: 57 cm/H: 47 cm/D: 52 cm (W: 22.4''/H: 18.6''/D: 20.3'')
Min. space required
(feet in forward position)
Size of the door
W: 55 cm/H: 42 cm/D: 38 cm (W: 21.5''/H: 16.4''/D: 15.6'')
movement
54 kg (119 lbs)
Weight empty
Max. weight (fully loaded) 71.5 kg (157.6 lbs)
Weight per support area
42 kN/m
2
Environment pollution
Degree 2
Usage environment
Indoor
STERILIZER CHAMBER
Pressure safety valve
2.6 bar (37.7 psi)
Safety thermostats
180 °C (356 °F)
~28 l, Ø: 279.5 mm/D: 456.8 mm ~7.4 gal, Ø: 11''/D: 18''
Total volume
Usable space *
W: 230 mm/H: 230 mm/D:390 mm (W: 9.05"/H: 9.05"/D: 15.35")
Bacteriological filter
0.3 µm
*: usable space with standard rack and trays. With optional racks and trays,
see Accessories, spare parts, consumables.
Technical data