DRAINING THE USED AND CLEAN WATER TANKS
Notice:
Before carrying out the following steps, make sure that the sterilizer
has completed the cycle in progress.
Always wear appropriate PPE during use of the sterilizer (e.g. gloves
for cleaning, maintenance, etc...).
 
If the used water tank is almost full, an alert appears on the
1
display.
 
Open the service door at the front of the sterilizer.
2
 
Put a container (5.5 l (1.45 gal)
3
min) below the sterilizer and insert
into it the free end of the drain tube.
 
Plug the other end of the drain tube
4
into the drain connector until it
clicks. Make sure that the connector
is firmly engaged.
Use left connector (blue, 1) for the
clean water and the right connector
(grey, 2) for the used water.
 
Let the water flow from the tank
5
completely.
22
 
Press the push-button on top of the
6
quick connector to dislodge the
drain tube.