REPLACE THE CONSUMABLE BEFORE THE MAINTENANCE DUE
DATE
If you replace the consumables before the request of replacement
appears, you should manually reset the counters through the
following procedure.
 
On the homepage, tap
1
 
Select the consumable you want to replace: a message appears
2
showing the current worked hours of the part.
 
Tap
to see an animated replacement procedure.
3
 
When you have replaced the consumable tap
4
executed cycle counter is reset.
Lara XL  | Instructions for Use |  RIS-311 ENG Rev02 |  01/06/2022 |  © W&H Sterilization Srl
>
to confirm: the
Maintenance