W&H Lara RIS-311 Instructions For Use Manual page 46

Hide thumbs Also See for Lara RIS-311:
Table of Contents

Advertisement

Sterilizer tests
 
Place the tube with the capsule in the center of a tray in the lowest rack position
4
and close the chamber door.
 
On the homepage, tap Helix-B&D test.
5
 
To set the duration of the Plateau/Sterilization phase and other settings, tap
6
 
Tap
and enter your credentials if required: the chamber door locks.
7
 
Wait until the end of the test and tap OPEN: the chamber door unlocks.
8
 
Enter your credentials if required.
9
 
Open the chamber door, extract the tray using the tray holder and take the tube.
10
CAUTION! Risk of burns. The test pack is very hot at the end of the cycle. Wear
appropriate PPE (e.g. gloves).
 
Unscrew the tube capsule and remove the indicator strip.
11
 
Check the change in colour. See "Interpret the test result" on the next page
12
46
.
Lara XL  | Instructions for Use |  RIS-311 ENG Rev02 |  01/06/2022 |  © W&H Sterilization Srl
 
 