4000 cycle or five-year maintenance

GENERAL CHECK AND SERVICE REQUIRED
Notice: Regular service is imperative to ensure continuous and
effective operation of the sterilizer.
A general check and service should be carried out every 4000 cycles
or five years by an authorized service technician. The service
required includes the following:
 
the replacement of consumables and other important internal
n
components
 
a check of the entire sterilizer with special care for the safety
n
systems
 
the cleaning of areas and components that cannot be accessed
n
by the user.
ACTIONS REQUIRED FOR EACH ELEMENT
For each element, the actions to carried out are the following:
Element
Electro valves
Vacuum pump internal parts
Sterilization chamber and external surfaces
Chamber filters
Lara XL  | Instructions for Use |  RIS-311 ENG Rev02 |  01/06/2022 |  © W&H Sterilization Srl
Replace
Clean
Check
x
-
-
x
-
-
-
x
-
-
x
-
Element
Internal parts, with particular care for the
condenser fins and the main board
Pneumatic connections
Electrical connections
Temperature and pressure calibration
Door locking system
Pressure safety valve
Safety systems
Maintenance
Replace
Clean
Check
-
x
-
-
-
x
-
-
x
-
-
x
-
-
x
-
-
x
-
-
x
75