MESSAGES AND ALERTS
Notice: For any error not listed in this table, call technical service.
Message/Alert
Fill clean water tank.
Drain used water tank.
Please close the door.
Non-conform water
Unacceptable water
Door Gasket must be replaced in
... cycles.
Have you ordered the Door
Gasket?
Bact. Filter must be replaced in
... cycles.
Have you ordered the Bact.
filter?
4000 cycle maintenance must
be performed in ... cycles.
Have you already booked the
4000 cycle service?
Lara XL  | Instructions for Use |  RIS-311 ENG Rev02 |  01/06/2022 |  © W&H Sterilization Srl
Description
There is not enough water in the tank to perform a
cycle.
The used water tank is full.
The door must be locked, but you didn't close it.
The clean water quality is bad (conductivity
between 15 and 50 μS/cm).
The clean water quality is very bad (conductivity
more than 50 μS/cm).
These are pre-alerts advising that one of the
consumables has to be replaced within a small
number of cycles.
This pre-alert advises you that the target of 4000
cycles is close and the relevant maintenance step
is to be scheduled.
Fill the water tank as requested.
Drain the water tank as requested.
Close the door so it can be locked.
You may run a cycle but the water must be replaced soon, otherwise the unit will
automatically lock-out to prevent damage.
Running a cycle is inhibited to prevent damage. Replace the clean water.
Tap
if you have the consumable available for replacement.
Tap
if you don't have the consumable in stock and must order one. In this case, the
pre-alert will appear again after some cycles.
See "Maintenance" on page  63.
Contact your Technical Service.
Action
Diagnostics
81