W&H Lara RIS-311 Instructions For Use Manual page 80

Hide thumbs Also See for Lara RIS-311:
Table of Contents

Advertisement

Diagnostics
Code
3xx
Check the door gasket. Clean or replace it if necessary.
Clean the chamber face side.
Check the load does not exceed the MAXIMUM WEIGHT LIMITS.
4xx
Clean water error (bad quality, clean tank low level, high consumption of water).
Drain and/or refill the clean water tank.
Check the door gasket. Clean or replace it if necessary.
Check if there are hurdles on the door locking area (chamber rack, loads, objects, ...).
5xx
Check the door gasket (wrong placed).
Check if the door can move freely without touching the trays or the load when closing.
Switch the sterilizer OFF and ON.
990
The cycle has been aborted by the user.
80
Description
Repeat the cycle.
If the problem persists, call service.
Repeat the cycle.
If the problem persists, call service.
Repeat the cycle.
If the problem persists, call service.
Re-process the load.
Lara XL  | Instructions for Use |  RIS-311 ENG Rev02 |  01/06/2022 |  © W&H Sterilization Srl
Actions

Advertisement

Table of Contents
loading

This manual is also suitable for:

Lara xlRis-311

Table of Contents