Technical data
STEAM GENERATOR
5 bar (72.51 psi)
Pressure safety valve
Safety thermostats
230 °C (446 °F)
DISTILLED OR DEMINERALIZED WATER
Fulfilling EN 13060 Ann. C (conductivity: < 15μS/cm, Total
Water quality
Dissolved Solids: < 10 ppm)
Average water
0.3 to 0.95 litres/cycle (0.63 to 2 gal/cycle)
consumption
Tank volume
Clean water 4.8 l (1.27 gal)
Min. water charge 0.82 l (0.22 gal)
Used water 5.11 l (1.35 gal)
100
Recommendations for validation
TEST VALIDATION POINTS
Lara XL sterilizers can be validated in accordance to EN ISO 17665-1.
For further details please refer to the Qualification/Validation guide
for sterilization cycles of the manufacturer.
Lara XL | Instructions for Use | RIS-311 ENG Rev02 | 01/06/2022 | © W&H Sterilization Srl
Part
Description
Hottest points
Coldest points