On the homepage, tap Helix-B&D test.
3
 
To set the duration of the Plateau/Sterilization phase and other settings, tap
4
 
Tap
and enter your credentials if required: the chamber door locks.
5
 
Wait until the end of the test and tap OPEN: the chamber door unlocks.
6
 
Enter your credentials if required.
7
 
Open the chamber door, extract the tray using the tray holder and take the test
8
pack. The test pack can be wet outside.
CAUTION! Risk of burns. The test pack is very hot at the end of the cycle. Wear
appropriate PPE (e.g. gloves).
 
Remove the indicator sheet from the center of the test pack and check the change
9
in colour. See "Interpret the test result" on the next page
Lara XL  | Instructions for Use |  RIS-311 ENG Rev02 |  01/06/2022 |  © W&H Sterilization Srl
.
Sterilizer tests
 
 
43