Maintenance
Monthly or 50-cycle maintenance
CLEANING THE DOOR GASKET AND THE CHAMBER FACE SIDE
Proceed as follows:
Clean the gasket seat and the chamber face side with a damp, lint-free cloth
1
moistened with clean water.
Notice:
Do not use abrasive products, cutting tools or sharp objects.
Rinse with clean water.
2
Note: when the seal is new it might be necessary to hold the door gently closed at the
sterilization start.
66
Lara XL | Instructions for Use | RIS-311 ENG Rev02 | 01/06/2022 | © W&H Sterilization Srl