Technical data
STANDARD STERILIZATION CYCLES AVAILABLE
There are four sterilization cycles available, all of them complying with the European Standard EN13060:
 
three B-type cycles
n
 
one S-type cycle (optional, activated with Fast Cycle cycle activation code. See "S Fast 134 cycle activation" on page  53)
n
Cycle
Cycle name
Purpose
type
B
B Universal 134
For all your general items like hand instruments, handpieces, forceps, etc.
Features an extended sterilization time, if this is required for your load or mandated in your country.
B Prion /
Extended 134 *
B Universal 121
For all items that cannot withstand the high temperatures of the 134 cycles, such as textiles and plastics.
S
S Fast 134
For fast processing of unwrapped instruments only, including dental turbines and handpieces, solid and hollow B (simple hollow items). It is not
(optional)
suitable for textiles, porous or bagged/wrapped items.
The instruments sterilized with this cycle cannot be stored: they must be used immediately after being sterilized.
*: cycle name might be different, depending on country requirements.