No function after starting
- Compressed air not connected yet
- Compressor not in use
- Air supply broken
- Air pressure too low
- Flap valve (02686800) defective
Cycle time more/less than 35±5 sec.
- Flap valve (02686800)
dirty / defective
- Timer (no. 4) dirty / defective
12
3
6
No rotation on the instrument
- Resistance of the instrument
too high
- Drive (no. 1) defective
- Tube to the drive detached
Air exit noises in the unit
- Crack/Hole in a tube
- Tube detached of the connection
- Pressure regulator soiled
Level of liquid indicator drops
too fast
- Tubing between tank and liquid
indicator defective
- Leakage of a tank
W&H Dentalwerk Bürmoos Ges.m.b.H., A - 5111 Bürmoos / Austria
Die angeführten Bauteile und
Nummern finden Sie in den
Zeichnungen " Ersatzteile und
Reparatur ASSISTINA 301 plus" ,
Nr. et00030140.
repair procedure
(no.)
1
2
3
4
5/6
5/6
5/6
7
18/5/6
8
6
8
5/6
6
6
mögliche
Problembehebung
Ursache
Keine Funktion nach dem Starten
- Luft noch nicht angeschlossen
- Kompressor nicht in Betrieb
- Luftzuleitung unterbrochen
- Zu geringer Antriebsdruck
- Rückschlagventil (02686800)
defekt
Zykluszeit weit über/unter 35±5 sek.
- Rückschlagventil (02686800)
verschmutzt / defekt
- Zeitglied (Nr. 4) verschmutzt /
defekt
Keine Rotation am Instrument
- Zu hoher Widerstand des
Instrumentes
- Antrieb (Nr. 1) defekt
- Schlauch zum Antrieb hat sich
gelöst
Luftaustrittsgeräusche im Gerät
- Riß/Loch in einem Schlauch
- Schlauch hat sich vom Anschluß
gelöst
- Druckregler verschmutzt
Pegel der Füllstandsanzeige fällt
zu schnell
- Schlauch zwischen Tank und
Füllstandsanzeige defekt
- Undichtheit eines Tanks
(Nr.)
1
2
3
4
5/6
5/6
5/6
7
18/5/6
8
6
8
5/6
6
6
1 / 6