Maintenance
CLEAN THE WATER TANK
Notice: Do not touch the water level sensors. If misplaced or misaligned from their original position, the operation of the sterilizer could be
impaired.
Proceed as follows:
 
Clean the internal tank surfaces with a soft sponge moistened with water. For
1
better cleaning results, clean with W&H MC-1000 cleaning solution. Use a small
non-abrasive brush to clean the areas that are difficult to reach.
Notice: never use any other disinfectant, detergent or abrasive product, as they
might result aggressive for the tank material.
 
Rinse the internal tank surfaces throughly, until all residuals of dirt and detergent
2
have been eliminated.
 
Dry the internal tank surfaces.
3
 
Remove the internal filter.
5
 
Clean the metal cartridge of the internal filter with tap water.
6
 
Reposition the internal filter.
7
 
Detach the drain tube.
8
72
Lara XL  | Instructions for Use |  RIS-311 ENG Rev02 |  01/06/2022 |  © W&H Sterilization Srl