END OF A STERILIZATION CYCLE
When a cycle is successfully finished, the "Cycle completed"
message appears on the screen. To end the cycle:
 
Tap
to view the cycle summary or tap
1
parameters. See "View the cycle parameters" on the previous
page.
 
Tap OPEN to open the door: the door unlocks and the home
2
page appears.
Note: if an error message appears see "Troubleshooting" on page  83
CAUTION! Hot surfaces.
Burnings. Do not touch the
chamber, the internal side
of the door and the internal
fittings. Use the tray holder
or cassette holder or
gloves for high
temperatures or adequate
protection to remove the
load!
 
Open the chamber door.
3
 
Remove the load and stock it.
4
Lara XL  | Instructions for Use |  RIS-311 ENG Rev02 |  01/06/2022 |  © W&H Sterilization Srl
STOP A STERILIZATION CYCLE
to view the cycle
A cycle can be manually aborted at any time. To stop a cycle:
 
 
 
 
Notice: Do not switch off the sterilizer during the reset phase: it
takes some time to reset the system and reach safe conditions in
the sterilizer chamber.
 
 
 
 
WARNING! You can stop the cycle at any time. Instruments
must not be considered sterile if this occurs before the
DRY phase.
Tap STOP: a confirmation request appears.
1
Tap
to abort the stop command. The cycle continues
2
programmed.
Tap
to abort the cycle: the sterilizer starts a reset phase.
3
Check the message. See "Messages of a stopped sterilization
4
cycle" on the next page.
Tap
to view the cycle parameters. See "View the cycle
5
parameters" on the previous page.
Open the chamber door.
6
Reprocess the load if necessary.
7
CAUTION! Hot steam. Wait the steam to dissipate before
opening the door.
Sterilization cycles