Installing The Sterilizer - W&H Lara RIS-311 Instructions For Use Manual

Hide thumbs Also See for Lara RIS-311:
Table of Contents

Advertisement

Getting started

Installing the sterilizer

LOCATION REQUIREMENTS
Notice:
Do not place the sterilizer so that it is difficult to operate the controls
behind the service door. Do not place the sterilizer so that it is
difficult to disconnect the power cord.
Leave the condenser grid (rear side of the sterilizer) free from
anything that might obstruct the air passage.
Surface materials should be water resistant. If sterilization cycles
will be continuous, pay attention to the surrounding materials: steam
can damage them.
The sterilizer must operate in absence of explosive atmospheres.
The sterilizer must operate in a well ventilated room (indoor), far
from sources of heat and from flammable materials.
Place the sterilizer on a flat and level surface.
20
Clearance requirements to ensure proper air circulation:
Make sure that the clean water tank can be easily accessed for
periodic cleaning operations.
ELECTRICAL CONNECTIONS
All the cables and tubes connected on the rear side of the sterilizer
must be placed far from the condenser grid (e.g. using the available
guides).
Notice:
Connect the sterilizer to a dedicated line. Do not use cable
extensions nor multiple sockets/adapters.
Ensure that external and internal surfaces are free from moisture or
condensation before connecting to power.
The installation of the sterilizer shall be performed by two authorized
technicians using PPE (Personal Protective Equipment) according to
applicable standards.
Lara XL  | Instructions for Use |  RIS-311 ENG Rev02 |  01/06/2022 |  © W&H Sterilization Srl