Sterilization cycles
DENTAL HANDPIECE EXTERNAL CLEANING
This procedure involves the removal of residues (blood, dentine,
etc.) that adhere to critical areas such as spray outlets, light ports,
knurling etc.
 
Wear protective gloves during cleaning.
n
 
Refer to the instructions of the instrument manufacturer.
n
 
Use a soft, damp brush and take care not to scratch the surface
n
of the light ports.
DENTAL HANDPIECE LUBRICATION
Once the dental handpieces has been disinfected, cleaned and dried
(free from residues), it must be lubricated prior to sterilization.
Follow manufacturer's instructions for proper lubrication.
PACKAGING
In order to preserve sterility, rotating instruments should be
wrapped/bagged prior to sterilization. Follow the manufacturer's
packing instructions when using sterilization packaging.
CLEANING THE INSTRUMENTS
Clean all instruments thoroughly prior to sterilization. If possible,
clean instruments immediately after use; always follow the
instrument manufacturer 's instructions. Remove all traces of
disinfectants and detergents. Rinse and dry carefully all
instruments.
The instruments and tubes must be carefully rinsed and dried prior
to sterilization.
50
CORRECT LOAD PLACEMENT
WARNING! Do not overload trays and the chamber. Adhere
to the maximum load weight limits (see "Sterilization
programs" on page  1).
Never place the load or the trays directly into the chamber
without the chamber rack as this could affect the steam
and temperature distribution. The load must always be
supported by the chamber rack.
Risk of burns. Before touching, ensure the sterilization
chamber is cold.
Wrap items with porous wrapping materials to facilitate steam
penetration and drying (e.g. sterilization bags for autoclaves).
Always use the chamber rack to allow adequate steam circulation.
Follow these requirements:
Load type
Placement
Hinged instruments (e.g.,
In open position
forceps, extraction pliers, etc.)
Tubes
Place tubes on a tray allowing the ends to
remain open. Do not bend tubes.
Cassettes can be placed vertically or horizontally
Cassettes
into the chamber rack (vertical placement
enhances drying). When placing cassettes
horizontally, slide them into the rack position
without putting them on trays (if size allows) to
enhance drying. When sterilizing double-decker
cassettes, place them in the lowest rack position
as there is more space height-wise.
Lara XL  | Instructions for Use |  RIS-311 ENG Rev02 |  01/06/2022 |  © W&H Sterilization Srl