Sterilization cycles
SET THE DRYING TIME
For each ECO DRY mode, the drying time is automatically adjusted to
the total amount of load and cannot be modified. To set a new drying
time for the program, you should disabled the ECO DRY mode first.
On the homepage, tap the cycle and
1
Select Fixed to disable the ECO DRY mode.
2
Tap the wording Fixed to modify the fixed mode minutes.
3
Tap
or
to increase the minutes or decrease them.
4
Note: for the minimum value of the drying time for each cycle see
"Sterilization cycles" on page 91.
Tap
to confirm and go back to the previous page. This
5
becomes the new fixed value.
CHANGE THE ECO DRY MODE
On the homepage, tap the cycle and
1
Tap the leaf icon to set the ECO DRY mode.
2
Tap the ECO DRY mode you desire.
3
Tap
to confirm and go back to the previous page.
4
Tap
to go back to the homepage.
5
56
.
.
VIEW THE CYCLE PARAMETERS
You can check the real time cycle parameters or the cycle
parameters at the end of the cycle. Following is an example:
While the sterilization cycle is running or when cycle ends tap
1
: the cycle information page opens.
Tap
or
to scroll the pages.
2
STERILIZATION CYCLE PAGE
Following are the information displayed while a cycle is running:
Lara XL | Instructions for Use | RIS-311 ENG Rev02 | 01/06/2022 | © W&H Sterilization Srl
Part
Description
Sterilization cycle name
A
ECO DRY mode enabled
B
C
Countdown clock (time until
the cycle completion)
D
: chamber pressure
: cycle counter
: chamber temperature
Date and time and door
E
securely locked symbol
Current cycle phase
F
Stop button
G
H
Button to open the cycle
information page