Pre-Cleaning (performed at bedside immediately after use of Ultrasonic Endoscope)

The pre-cleaning (the first cleaning) is performed at the bedside immediately after the use
of the ultrasonic endoscope.
6.7.1

Wiping

Wipe contamination off the outside of insertion portion of
Ultrasonic Endoscope with gauze or paper towel.
6.7.2

Cleaning Suction Channel

You are going to clean suction channel.
(1) Put distal end of Ultrasonic Endoscope in cleaning fluid,
and depress suction button to suck in cleaning fluid for 10
seconds.
(2)
Keeping suction button depressed, pull distal end of
Ultrasonic Endoscope out of cleaning fluid and suck in air.
(3) Repeat steps (1) and (2) above 2 or 3 times to suck in
cleaning fluid and air alternately.
(4)
Finally, pull distal end of Ultrasonic Endoscope out
of cleaning fluid and suck in air until cleaning fluid is
completely drained from suction channel.