6.3 Cleaning the Air Intakes and Exhaust Ports

Clean air filter, louvers and air intakes of the right-side cover and the exhaust ports of the left-side
cover periodically (once every three months).
1
Press the button on the top of the louver of
the right side (air intake) (1.) and dismount
the louver (2.).
1.
Press
2
Remove air filter attached to the inside of the
upper louver.
No air filter is attached to the inside of the lower
louver.
6
Remove
3
Use a vacuum cleaner to clean the front and
back of the air filter, louvers and air intakes.
4
Reattach the air filter to the upper louver.
Reattach
6-4
Upper louver
2.
Dismount
Lower louver
2.
Dismount
Air filter
Air filter
5
Remount the louvers on the equipment.
Remount
Remount
6
Use a vacuum cleaner to clean the exhaust
ports of the left side.
When remounting the louver, be careful not to
pinch your fingers.
FCR XC-1 Operation Manual
Upper louver
Lower louver
Exhaust ports