Default Values
Verification kits: Verikit low: ..................................................... 0.0900
Verification kits: Verikit high: .................................................... 0.1800
Filter Time Const.: ........................................................................30 s
Hold after Cal.:............................................................................120 s
Meas. Interval: ............................................................................6 min
.........................................................................same as in Installation
.........................................................................same as in Installation
.........................................................................same as in Installation
Logger Interval:............................................................... event driven
Clear Logger:.................................................................................. no
Disinf:........................................................................................Ozone
Reag. expiry warning.....................................................................yes
Parameter:................................................................................... DIS.
Current loop:........................................................................ 4–20 mA
Function:.....................................................................................linear
Scaling: Range low:................................................................ 0.0 ppb
Scaling: Range high: .......................................................... 500.0 ppb
Parameter:.......................................................................Sample flow
Current loop:........................................................................ 4–20 mA
Function:.....................................................................................linear
Scaling: Range low:.................................................................... 0 B/s
Scaling: Range high: .............................................................. 200 B/s
Alarm DIS.:
Alarm high: ......................................................................... 500.0 ppb
Alarm low:............................................................................... 0.0 ppb
Hysteresis:.............................................................................. 1.0 ppb
Delay: .............................................................................................5 s
Sample Flow: Flow Alarm: .............................................................yes
Sample Flow: Alarm High: ...................................................... 500 B/s
Sample Flow: Alarm Low: ........................................................... 5 B/s
Case temp. high: ....................................................................... 65 °C
Case temp. low:........................................................................... 0 °C
Function:............................................................................Limit upper
Parameter:................................................................................... DIS.
Setpoint: ............................................................................. 500.0 ppb
A-96.250.881 / 070622