AMI Codes-II O3
Maintenance
6.3.1
Prepare
Oxycon
On-line DPD
Prepare
Oxycon
On-line Buffer
O3
Start-up
A-96.250.881 / 070622
Reagents for measuring Ozone
1 Rinse the canister [E] labelled OXYCON ON LINE DPD Reagent
with demineralized water.
2 Fill the canister up to the 2 liter mark with demineralized water.
3 Slowly pour the content of a bottle of concentrate Oxycon On-line
DPD into the canister.
Avoid splashing!
4 Close the canister with the screw cover and tighten it well.
5 Mix the demineralized water and the reagents well.
6 Put the canister [E] into the holder [G].
7 Remove the screw cover, insert the suction lance [A] and tighten
the screw cover.
1 Rinse the canister [F] labelled OXYCON ON LINE Ozone Re-
agent with demineralized water.
2 Fill the canister up to the 2 liter mark with demineralized water.
3 Slowly pour the content of one bag of Oxycon On-line Buffer 03
into the canister.
4 Close the canister with the screw cover and tighten it well.
5 Mix the demineralized water and the reagents well.
6 Put the canister [F] into the holder [G].
7 Remove the screw cover, insert the suction lance [B] and tighten
the screw cover.
1 Fill reagent system. See
2 Reset the reagent expiry counter in menu 3.2.3.3
<Maintenance>\<Service>\<Reagents>\<Change reagents>\
<yes>.
Fill or Flush Reagent System, p. 35.
45