AMI Codes-II O3
Maintenance
6.9.
56

Longer Stop of Operation

1 Put the suction lances into a bucket with demineralized water.
2 Start <Fill system>.
 The reagent tubes are flushed with water.
3 Remove the suction lance from the water.
4 Start <Fill system> again.
 The water will be pumped out of the reagent tubes.
5 Stop sample flow.
6 Wait until flow cell is empty.
7 Shut off power of the instrument.
8 Open the occlusion frames of the peristaltic pump, see
the Pump Tubes, p.
54.
Replace
A-96.250.881 / 070622