AMI Codes-II O3
Troubleshooting
7.2.
7.2.1
Possible error
message
7.3.
Possible error
message
58
Calibration Errors
Process Calibration DIS
Slope error:
Possible cause
Wrong manual measure-
ment.
Wrong reagent mixture
Reagents not
completely solved in
water.
Reagents older than 30
days.
Verification Errors
Invalid zero measurement:
Possible cause
Gas bubble in the pho-
tometer.
Verification error
Possible cause
Deviation between mea-
sured value and refer-
ence value is too high
Corrective Action
Repeat the manual measurement.
Use fresh reagents.
Make a correct mixture.
Mix long and intensively.
Corrective Action
Rinse the photometer with sample for
1-2 minutes and repeat the verifica-
tion.
Corrective Action
Clean the verification filters
according to
Cleaning, p. 48
repeat the verification.
If filter cleaning does not result in a
successful verification, clean and
rinse the photometer according to
Cleaning the Photometer, p. 51
repeat the verification.
If the problem perists, call service.