AMI Codes-II O3
Troubleshooting
Error
E021
DIS. invalid
E022
Reagent empty
E024
Input active
E026
IC LM75
E028
Signal output open
E030
EEprom Frontend
E031
Calibration Recout
E032
Wrong Frontend
E033
Power-on
E034
Power-down
E065
DPD / Buffer
E066
Reagents expired
A-96.250.881 / 070622
Description
Corrective action
– This error appears after start-up and will
disappear after the first valid
measurement is finished.
– refill reagents, see
Reagents, p. 43
– No action necessary.
– This message is displayed if "Fault =
Yes" is programmed, see
– call service
– check wiring on signal outputs 1 and 2
– call service
– call service
– call service
– none, normal status
– none, normal status
– Operating display, upper status line. The
number next to the triangle, indicates
the remaining reagents in%. Refill
reagents on time. See
Reagents, p. 43
– Canister filling is older than 30 days.
– This message is displayed if "Reag.
expiry warning = Yes" is programmed,
see
5.1.2, p.
73.
Refill or replace
5.3.4, p.
81.
Refill or replace
61