Maintenance
6.2.
42
Stop of Operation for Maintenance
1 Put the suction lances into a bucket with demineralized water.
2 Start <Fill system>.
The reagent tubes are flushed with water.
3 Remove the suction lances from the water.
4 Start <Fill system> again.
The water will be pumped out of the reagent tubes.
5 Stop sample flow.
6 Wait until level in flow cell is empty.
7 Shut off power of the instrument.