Replacing The Tungsten Lamp - Thermo Scientific Surveyor UV Plus Hardware Manual

Table of Contents

Advertisement

Installing a New Deuterium Lamp

Replacing the Tungsten Lamp

Thermo Scientific
To install a new deuterium lamp
1. Remove the used deuterium lamp as described in
page
40.
2. Hold the deuterium lamp assembly so that the leads are at the top.
3. Slide the assembly onto the alignment pin (see
Figure 32. Deuterium lamp assembly alignment pin
4. Securely fasten the assembly in place with the two thumbscrews and standoffs.
5. Connect the lamp lead to the upper terminal in the lamp compartment.
6. Replace the lamp housing and secure it with the screw.
7. Close the front doors of the detector.
8. Plug the power cord back into the power receptacle on the back panel of the detector.
To replace the tungsten lamp, follow these steps:
1.
Removing the Tungsten Lamp
2.
Installing a New Tungsten Lamp
"Removing the Deuterium Lamp"
Figure
32).
Alignment pin
Surveyor UV/Vis Plus Detector Hardware Manual
3
Maintenance
Lamp Maintenance
on
41