Surveyor Firmware Upgrade Utility
54
Surveyor UV/Vis Plus Detector Hardware Manual
6. In the Surveyor Firmware Upgrade Utility window, initiate communication between the
utility and the Surveyor UV/Vis Plus Detector as follows:
a. In the Device Type area, select the Surveyor UV/VIS option. See
CAUTION Do not interrupt the firmware download process. Do not turn off the
power to the Surveyor UV/Vis Plus Detector or close the Surveyor Firmware
Upgrade Utility while the detector is connected to the Surveyor Firmware
Upgrade Utility. If you lose power to the detector during the firmware download
process, contact your Thermo Fisher Scientific representative to reprogram the
firmware board.
Figure 36. Surveyor Firmware Upgrade Utility, showing the Surveyor UV/Vis option
selected
b. Click Connect.
The status display at the bottom of the Surveyor Firmware Upgrade Utility window
indicates that a connection has been made. See
Figure
Figure
37.