Standard Analytical Flow Cell

Removing the Standard Analytical Flow Cell Assembly

Thermo Scientific
To remove or install the standard analytical flow cell, follow these topics:
Removing the Standard Analytical Flow Cell Assembly
Installing the Standard Analytical Flow Cell
CAUTION Flow cells are factory-assembled units. Do not disassemble them.
To clean or replace the standard analytical flow cell, remove it from the UV/Vis detector.
To remove the standard analytical flow cell
1. Turn the detector off and unplug the detector from line power.
CAUTION To prevent possible exposure to UV light, turn the detector off and unplug
the power cord from the power receptacle on the back panel.
2. Open the front doors of the detector.
The flow cell assembly is located behind the flow cell cover.
3. Remove the flow cell cover by unscrewing the flow cell cover retention screw and pulling
it straight out and away from the flow cell assembly. See
Figure 25. Surveyor UV/Vis Plus Detector with the flow cell cover installed
Standard Analytical Flow Cell
Figure
25.
Standard analytical flow cell
cover retention screw
Surveyor UV/Vis Plus Detector Hardware Manual
3
Maintenance
33