4. Carefully pull the photodiode mount straight back and off the LightPipe mounting
assembly. Rest the photodiode mount out of the way inside the detector housing or on
the benchtop.
5. Loosen the thumbscrew holding the retention block to the LightPipe retention bar. See
Figure
21.
Figure 21. Retention block thumbscrew
Thumbscrew
6. Lift the retention bar shown in
the LightPipe flow cell.
Figure 22. Retention bar for the LightPipe flow cell
Retention bar
Retaining block
Notch
Figure 22
until the bar is clear of the alignment notch in
Optical fiber
Surveyor UV/Vis Plus Detector Hardware Manual
3
Maintenance
LightPipe Flow Cell
29