Installation
Making the Back Panel Connections
Connecting the 7-Connector System Interconnect Cable
18
Surveyor UV/Vis Plus Detector Hardware Manual
To connect the 7-connector system interconnect cable to the LC modules
1. To connect the Surveyor Autosampler Plus, plug the A/S connector into the left, 8-pin
socket on the back panel of the autosampler as shown in
2. To connect the pump, do the following:
• To connect the Surveyor LC Pump Plus, plug a connector labeled PUMP into the
8-pin socket on the back panel of the pump.
• To connect the Surveyor MS Pump Plus, do the following:
Connect the end of the adapter cable labeled PUMP to one of the PUMP
connectors of the 7-connector interconnect cable.
Plug the end of the adapter cable labeled MS PUMP into the 8-pin socket on the
back panel of the MS pump.
pump.
Figure 15. Adapter cable for the Surveyor MS Pump Plus
Connect to
the system interconnect cable
3. To connect the UV/Vis detector, plug a detector connector into the left, 8-pin socket on
the back panel of the detector.
Figure
Figure 15
shows the adapter cable for the MS
7 inches
(18 cm)
60053-63038
16.
Connect to
the pump