Thermo Scientific Surveyor UV Plus Hardware Manual page 55

Table of Contents

Advertisement

Thermo Scientific
2. Open the front doors of the detector.
CAUTION Because the lamps run hot, turn off the power and allow sufficient time
(approximately 30 minutes) for the lamps to cool before you remove the lamp cover.
3. Using the 1/8 in. ball-end hex screwdriver (supplied), unscrew the captive retaining screw
on the lamp cover (see
Figure 28. Lamp cover retaining screw
Lamp cover retaining screw
4. Pull the cover off to expose the lamp assemblies.
5. To read the chronometer, check the position of the electrolyte gap in the mercury tube
against the graduated background (see
Note
Figure
29,
Figure
deuterium lamp.
Figure 29. Chronometer for the deuterium lamp
Figure
28).
Figure 29
31,
Figure
33, and
Figure 35
Surveyor UV/Vis Plus Detector Hardware Manual
and
Figure
30).
show the latest version of the
3
Maintenance
Lamp Maintenance
39

Advertisement

Table of Contents
loading

This manual is also suitable for:

Surveyor vis plus

Table of Contents