Installing a New Tungsten Lamp
Thermo Scientific
Figure 34
shows the tungsten lamp assembly.
Figure 34. Tungsten lamp assembly
To install a new tungsten lamp
1. Remove the used tungsten lamp as described in
page
42.
2. Hold the lamp assembly so that the leads are at the top. Slide the assembly onto the two
alignment pins shown in
detector monochromator aperture.
Figure 35. Alignment pins for the tungsten lamp
"Removing the Tungsten Lamp "
Figure
35. The alignment pins are located on either side of the
Surveyor UV/Vis Plus Detector Hardware Manual
Lamp Maintenance
Tungsten bulb
Alignment holes
Alignment pins
3
Maintenance
on
43