Surveyor Firmware Upgrade Utility
56
Surveyor UV/Vis Plus Detector Hardware Manual
Figure 38. Surveyor Firmware Upgrade Utility, showing a successful download
9. Close the Surveyor Firmware Upgrade Utility. The connection with the Surveyor UV/Vis
Plus Detector automatically terminates.
10. Reset the Unit ID on the back panel of your detector for normal operation as follows:
a. Turn off the power to the UV/Vis detector.
IMPORTANT
b. Set the rotary switches back to their original positions.
c. Turn on the UV/Vis power to resume normal operation.
Turn off the UV/Vis power before adjusting the rotary switches.