Safety Barrier (E Ia); Cleaning Unit - Siemens MAG 1100 Operating Instructions Manual

Electromagnetic flowmeters
Table of Contents

Advertisement

SITRANS F M
2.5.4

Safety barrier (e ia)

2.5.5

Cleaning unit

2. Technical data
Application
For use with MAG 5000/6000 19" and MAG 1100 1100 F and 3100, 3100 P, in
ATEX 2 GD version
Ex approval
[EEx e ia] IIC
Cable parameter
Group
Electrode
IIC
IIB
IIA
During operation: −20 ... +60°C
Ambient temperature
During storage: −20 ... +70°C
Enclosure
Material
Standard 19" insert in aluminium/steel (DIN 41494)
Width: 21 TE
Height 3 HE
Enclosure rating
IP 20 to EN 60529 and DIN 40050
Mechanical load
1 G, 1-800 Hz sinusoidal in all directions to EN 60068-2-36
EMC performance
Emission
EN 50081-1 (Light industry)
Immunity
EN 50082-2 (Industry)
Application
For use together with MAG 5000 and 6000 19" insert to clean the measuring
electrodes on sensors MAG 1100, MAG 1100F or MAG 3100.
NB: Must not be used with intrinsically safe ATEX sensors
NB: Not be used with sensors with Hastelloy and Tantalum electrodes
NB: Only for measuring electrodes
Cleaning voltage
AC cleaning
60 V AC
DC cleaning
30 V DC
Cleaning period
60 sec. + 60 sec. pause
Relay
Switch relay activated when cleaning is in progress
Load
42 V/2 A
Operation
Switch relay activated when cleaning is in progress
Automatic
Yes
Manual
No
Indicator lamps
LEDs: "ON" and "CLEANING"
115 ... 230 V AC +10% ... −15%, 50 ... 60 Hz, 7 VA cleaning, 5 VA stand by
Supply voltage and
power consumption
11 ... 30 V DC /11 ... 24 V AC, 50 ... 60 Hz, 7 VA cleaning, 5 VA stand by
During operation: −20 ... +50°C (−4 ... +122 °F)
Ambient temperature
During storage: −20 ... +70°C (−4 ... +158 °F)
Enclosure
Material
Standard 19" insert in aluminium/steel (DIN 41494)
Width:
21 TE
Height:
3 HE
Rating
IP 20 to EN 60529 and DIN 40050
Mechanical load
1 G, 1 ... 800 Hz sinusoidal in all directions to EN 60068-2-36
Cleaning unit
The Siemens Flow Instruments cleaning unit can be used with MAG 5000 or 6000 in 19" insert non-
CT version.
The cleaning unit can be used in applications where the liner material and subsequently the
electrodes may be coated with deposits. If the coating is electrically insulating, the electrode signal
will be reduced. If the coating is electrically conductive, the electrode signal will be partly short-
circuited and in both cases the accuracy of the meter will decrease (dependent on the type and
thickness of the coating).
Note:
The cleaning unit cannot be used for flammable or explosive media!
Note:
The cleaning unit cannot be used with sensors with Hastelloy and Tantalum electrodes!
Safety barrier & cleaning unit
Capacity in μF
≤ 4.1
≤ 45
≤ 45
Inductance in mH
≤ 1.5
≤ 87
≤ 87
19