P052 Tank Dimension A [MR 200]
Dimension A as used in P050 Tank Shape.
Primary Index
Values
Related
Enter one of the following:
height of the tank bottom if P050 = 2,3,4, or 5
OR
length of one end section of the tank if P050 = 7, in Units (P005)
P053 Tank Dimension L [MR 200]
Dimension L as used in P050 Tank Shape.
Primary Index
Values
Related
Enter the following:
tank length (excluding both end sections) if P050 = 7
P054 Level Breakpoints (Universal Volume Calculation)
[MR 200]
When the tank shape is too complex for any of the preconfigured shapes, you can specify
the volume based on segments. See "Accessing a Secondary Index" on page 117 for
more information
Primary Index
Secondary Index
Values
Related
Enter the following:
up to 32 level breakpoints (where volume is known) if P050 = 9 or 10
Page 126
MultiRanger 100/200 – INSTRUCTION MANUAL
Level
Range: 0.0 to 9999
Preset: 0.000
P050 Tank Shape
Level
Range: 0.0 to 9999
Preset: 0.000
P050 Tank Shape
.
Level
Breakpoint
Range: 0.0 to 9999
P055 Volume Breakpoints
7ML19981FB05