Enter the factor (powers of 10 only) by which actual volume is divided, prior to Remote
Totalizer count increment.
Example:
For a Remote Totalizer update by 1000s of volume units, enter 3.
P641 Flow Sampler Mantissa [MR 200]
Use this feature in conjunction with Flow Sampler Exponent (P642) to establish the
number of flow units required to increment the Flow Sampler (device connected to the
MultiRanger relay set for the flow sampler operation Relay Function, P111 = 41).
Primary Index
Values
Related
This parameter is relevant only if Operation is set to OCM (P001 = 6).
Enter the mantissa (Y) for the exponent (Z) in the formula:
Flow Sampler Increment = Y x 10
To count once every 4310 (4.31 x 10
Example:
set P641 to 4.31 and P642 to 3
P642 Flow Sampler Exponent [MR 200]
Use this feature in conjunction with Flow Sampler Mantissa (P641) to establish the
number of flow units required to increment the Flow Sampler (device connected to the
MultiRanger relay set for the flow sampler operation Relay Function, P111 = 41).
Primary Index
Values
Related
7ML19981FB05
MultiRanger 100/200 – INSTRUCTION MANUAL
Single Point Model
Global
Range: 0.001 to 9999
Preset = 1.000
P001 Operation
P111 Relay Control Function
OCM (P600 to P621)
P642 Flow Sampler Exponent
Z
Flow units.
3
Single Point Model
Global
Range: -3 to +7 (integers only)
Preset = 0
P001 Operation
P111 Relay Control Function
OCM (P600 to P621)
P641 Flow Sampler Mantissa
Dual Point Model
Transducer
) flow units:
Dual Point Model
Transducer
Page 177